Reliable & Cost-Effective Mercury Emissions Sampling – EPA Method 30B Compliant
Overview
The XC-260 Manual Mercury Sampling Console is a low-cost, portable, and field-proven solution for sampling mercury emissions compliant with CFR 40, Part 60, Method 30B.
At the heart of this system is the XC-260 Source Sampler, a precision meter console designed for accurate mercury emissions collection. Mercury concentration is determined using sample volumes measured by the dry gas meters combined with sorbent trap analysis results.
Designed for Field Use
✔️ Lightweight & portable – Easy to transport for on-site mercury testing
✔️ Dual independent flow paths – Flexible operation for enhanced efficiency
✔️ Durable roto-moulded polyethylene case – Built for rugged field conditions
Key Features & Benefits
✔️ Dual dry gas meters with digital LCD displays for accurate measurement
✔️ Coarse & fine control valves for precise flow & vacuum adjustments
✔️ Stainless steel valves & fittings for maximum durability
✔️ 1/4-inch stainless steel quick-connects for secure, leak-free connections
✔️ Digital temperature controller for precise probe heater regulation
✔️ Elapsed digital timers for real-time monitoring
✔️ Dual-scale vacuum gauge for accurate system pressure readings
✔️ Multiple flow rate configurations available:
- 100–1000 cc/min
- 200–2400 cc/min
- 200–4000 cc/min
✔️ Available in 120V at 60Hz or 240V at 50Hz for global compatibility
Technical Specifications
- Dry Gas Meters:
- Model SK25EX series, positive displacement diaphragm type
- 0.7 litres per revolution
- Optical encoder sensor with quadrature pulse output
- Gas volume totalizer with 8-digit LCD display (1 cc resolution)
- Dual-Diaphragm Sample Pumps: (2) 12V DC motors, max vacuum 20 inHg
- Probe Temperature Control:
- Compact auto-tuning digital controller (1/32 DIN)
- External 25A solid-state relay
- 115V AC outlet (5-15R), standard Type-K thermocouple input
- Digital Temperature Display:
- 3.5-digit display (-157°F to 1999°F)
- Resettable for °C readings (°C is standard for exports)
- Thermocouple Inputs: 5 external Type-K inputs + 2 Type-K internal gas meter connections
- Flowmeter Options:
- A: 0.1 – 1.0 LPM
- B: 0.2 – 2.4 LPM
- C: 0.2 – 4.0 LPM
- Power Supply: 120V at 60Hz or 240V at 50Hz (optional)
- Dimensions: 23” W × 21” H × 12” D (58 cm × 53 cm × 30.5 cm)
- Weight: 39 lbs (17.7 kg)
Reliable, Cost-Effective & Easy to Transport
The XC-260 Mercury Sampling Console is the ideal solution for field testing mercury emissions, offering accuracy, durability, and ease of use.