Advanced Automation for Isokinetic Sampling – Method 5 & Beyond
Overview
The XC-5000 AutoKinetic™ Sampling Console is designed for US EPA Method 5 and associated isokinetic methods, bringing automation, precision, and efficiency to emissions sampling.
This state-of-the-art console eliminates manual data entry errors and complex calculations, ensuring high-accuracy measurements and streamlined reporting. Its bright LED display provides clear, real-time data visibility in any lighting condition, while AutoKinetic proprietary software offers a user-friendly, stepwise interface to guide operators through the sampling process.
Smart, Automated & Reliable
✔️ Automated sampling console improves data integrity & eliminates errors
✔️ Windows-based interface simplifies operation with guided test setup
✔️ Automatic calculations – traverse points, nozzle diameter, isokinetic rate
✔️ Pre- and post-leak checks with software assistance
✔️ Alerts & notifications – prompts users to move probes & confirm test segments
✔️ Built-in report preparation – downloadable data for easy documentation
✔️ Compatible with standard Method 5 stack sampling components
Key Features & Benefits
✔️ Rugged, lightweight roto-moulded polyethylene case with ergonomic handles
✔️ Multi-language support with English/metric unit selection
✔️ 7” full-colour, sunlight-readable LED display
✔️ Automatic pause function for flexible operation
✔️ Data storage & sampling profile management
✔️ Ethernet connectivity for seamless communication
✔️ Accepts standard modules for various EPA Methods
Technical Specifications
- Dry Gas Meter: Precision DGM, 0.7L per revolution, digital encoder (0.2L resolution)
- Temperature Control: Integrated system via control & data acquisition board
- Thermocouple Display: Seven simultaneous readings (probe, stack, oven, filter, exit, aux, DGM)
- Digital Pressure Transducers:
- Pitot ΔP: ±2.5 inH₂O (±63 mm), Resolution: 0.01 inH₂O (0.1 mm)
- Orifice ΔH: 0 – 5 inH₂O (0 – 127 mm), Resolution: 0.01 inH₂O (0.1 mm)
- Barometric Pressure: 17.7 – 32.5 inHg (450 – 825 mmHg), Resolution: 0.01 inHg (0.1 mmHg)
- Vacuum Sensor: 0 – 30 inHg (0 – 101 kPa), Accuracy: ±2%
- Umbilical Connections:
- Electrical: 4-conductor circular connector (grounded)
- Sample Line: Stainless steel 1/2-inch quick connect
- Pitot Line: Stainless steel 1/4-inch quick connects (optional 3/8”)
- External Pump: Stainless steel 3/8-inch quick connect
- Thermocouples: Type-K standard size
- Communication: Ethernet connectivity
- Optional Module: 4-channel analog input for external data logging (4-20mA, 0-10V, 1-5V)
- Power Supply: 120V at 60Hz or 240V at 50Hz (optional)
- Console Power Requirements: 15A max
- Dimensions: H23” × W21” × D12” (58 cm × 53 cm × 30.5 cm)
- Weight: 39 lbs (17.7 kg)
No More Manual Errors – Let the XC-5000 Handle the Hard Work
The XC-5000 AutoKinetic™ Sampling Console enhances accuracy, efficiency, and reliability, making it the ultimate solution for isokinetic sampling in compliance with EPA Method 5 and related methods.